Human Pituitary LINCOplex Kit

 All components are included in each kit (capture antibody immobilized on microspheres, bead diluent, detection antibody cocktail, standards, quality controls, streptavidin-PE,wash buffer, assay buffer, 96-well microtiter filter plate, matrix solution) and are sufficient to run 38 samples (serum, plasma or tissue culture medium) in duplicate in a 96-well microtiter plate.

This multiplex assay kit can be used for the simultaneous quantitative determination of the following six human pituitary hormones in any combination: FSH, LH, TSH, Prolactin, GH, and ACTH. LINCOplex® is especially useful for conducting multiplex assays on samples where limited volumes are available.

ASSAY CONDITIONS:  
25 µL sample volume
4 hour assay at room temperature

CROSS REACTIVITY:  
Each antibody pair used in this panel is highly specific and does not show any significant
cross-reactivity to other analytes within this panel.

STANDARD CURVE RANGE:  
Human Pituitary Hormones: Range
FSH           0.024-100 mIU/mL
LH             0.049-200 mIU/mL
TSH           0.012-50 µIU/mL
Prolactin     0.024-50 ng/mL
GH             0.012-50 ng/mL
ACTH      0.003-12.5 ng/mL
 

ACCURACY/RECOVERY
Human Pituitary Hormones
FSH 92.3%  
LH 89.5%  
TSH 113.2%  
Prolactin 91.6%  
GH 76.4%  
ACTH 90.6  

 

SENSITIVITY:
Human Pituitary                          
Horomones:                                    
MinDC*
FSH 0.005 mIU/mL
LH 0.007 mIU/mL
TSH 0.014 µIU/mL
Prolactin 0.004 ng/mL
GH 0.004 ng/mL
ACTH 0.003 ng/mL


MINIMUM DETECTABLE CONCENTRATIONS

PRECISION: Human Pituitary Hormones:
                       Intra-Assay      Inter-Assay
                 Precision (%CV)  Precision (%CV)
FSH                      7.2                    4.9
LH                        6.3                    6.7
TSH                      6.9                    3.0
Prolactin                7.0                    8.2
GH                        5.4                  14.9
ACTH               10.8                  7.4